Service and Repair
| Calibration |
| |
| DURRIDGE maintains a radon calibration facility that includes
a controlled, standard source of radon gas. All RAD7 alignment
and calibration is done here, as well as basic testing and
quality assurance.
We determine calibration factors by direct comparison to "master"
RAD7's, which were themselves compared with EPA and DOE instruments,
and which have participated in international inter-comparisons
of radon instrumentation. The calibration accuracy is independently
verified by direct determination of the radon chamber level
from the calibrated activity and emission of the standard radon
source. In addition, we regularly intercompare with other
radon chambers. We achieve a reproducibility of better
than 2% with our standard RAD7 calibration. Overall calibration
accuracy is better than ±5%. |
| |
| The EPA recommends, and we agree, that continuous radon
monitors be calibrated at least every year.
RAD7 calibration requires four to five working days from the
receipt of the
instrument. And we usually return it to you via the same shipping
method you used to get it to us - from "slow boat" to "next
day air". As a preliminary to recalibration, we inspect
and test every RAD7. If
additional service is
required,
this may delay the return
of the instrument. |
| |
| We recommend that you arrange for your calibration ahead
of time, to avoid possible delays. When sending your RAD7 for
calibration, please send the instrument only, without the accessories,
and please pack it well, with at least an inch of packing all
round. |
| |
| Thoron Calibration |
| |
| The RAD7 is able to measure both radon (222-Rn) and thoron
(220-Rn) concentrations simultaneously and independently. The
standard radon calibration is performed with a precision of
better than 1% and we guarantee the RAD7 absolute accuracy
for radon to be better than 5%. However thoron has a very short
half life (about one minute) so that the concentration in the
air will change during sample acquisition. For every minute
the sample gas is in transit from the sampling point to the
measurement chamber, the concentration of thoron in the sample
will drop by half. |
| |
| The RAD7 manual describes the standard setup for thoron sniffing.
It is designed to minimize the transition time of the air sample
from sampling point to measurement chamber. The pump runs continuously
and the desiccant is in a small tube. In those conditions,
the acquisition time is about one minute, so the default thoron
sensitivity of the RAD7 is set at half the radon sensitivity.
This setting is sufficient for most applications. |
| |
| For example, when trying to locate radon entry points, it
is sufficient to find locations where there is measurable thoron
to know that there is a fresh soil gas entry point nearby.
Anywhere else in the building, far from a radon entry point,
the thoron will have decayed away to practically nothing, so
just finding thoron and zeroing in on the highest reading achieves
the object of locating the radon entry point. For this application,
it is necessary to use the RAD7 in thoron protocol, and it
is necessary for the thoron readings to be repeatable and in
correct relation to each other. The standard (thoron protocol)
RAD7 setup, with the default thoron sensitivity setting, provides
all the functionality and accuracy necessary for this application. |
| |
| Occasionally a user requires more accurate thoron readings.
For these applications DURRIDGE Company can provide, as an
option, a specific thoron calibration for the setup used by
the user. The detailed conditions of the calibration, including
the air flow rate, may be chosen by the user and are clearly
specified in the thoron calibration certificate issued by DURRIDGE.
For any other setup, with a different flow rate or a different
volume in the sample air path, that thoron calibration is no
longer valid. For some projects, DURRIDGE Company has provided
multiple thoron calibrations, each for a different setup of
the RAD7 and air sample acquisition path. |
| |
| Please note that while the thoron sensitivity of the RAD7
is strongly dependent on the sample air transit time, from
sampling point to measurement chamber, the radon sensitivity
of the instrument is constant (to within 1%) and independent
of the sample air transit time provided this is less than one
hour. The standard radon measurement (not thoron protocol)
will refresh the air in the measurement chamber once every
five minutes, so that the radon level is kept up to date. |
| |
| Repair |
| |
| If you discover that your RAD7 is malfunctioning, we recommend
that you first call DURRIDGE and talk to a technician. A surprising
number of minor "disasters" can be avoided by long-distance
consultation. The next step, if consultation fails, usually
is to send your instrument in for evaluation and repair. Please
send any documentation of the problem that you might have (notes,
printouts, etc.) and a short note describing the problem. Be
sure that you put your name and telephone number on the note.
Within 48 hours of our receipt of the instrument, we will call
you to give a diagnosis. |
| |
| Bear in mind that a repaired instrument often requires complete
re-calibration, so a one-week turn-around may be the best we
can do. If you absolutely need an instrument during the repair
time we can sometimes arrange to send you a "loaner" instrument. |
| |
| Shipping |
| |
| The RAD7 is shipped to you in a plastic bag, packed in a
box with "popcorn", and finally packed inside another
box that also holds all the accessories. The inner box, by
itself,
is
too small for the return trip, when you need a calibration
or repair job. You need to find a bigger box, with space for
one inch packing all round the RAD7. A 14" cube box is
fine. Pack and seal the box carefully. |
| |
| Please do not ship any accessories, unless they are relevant
to the problem. |
| |
| All return shipments to DURRIDGE must be prepaid, and unless
you instruct us otherwise, we'll return it to you via the same
shipping method you used to get it to us - from "slow
boat"
to "next day air", freight collect. |
| |
| Upgrades |
| |
| Whenever you send your RAD7 in for repair or calibration,
you have the option of having the latest available software
installed. Most RAD7's can be upgraded to the latest hardware
configuration as well. You will be informed periodically of
whatever new features are available for your RAD7. Please advise
if you want to have an upgrade made. We intend to keep our
RAD7 customers happy by keeping their instruments up-to-date,
state-of-the-art. |
| |
|
| |
|
|
|
 |
|
 |
|