1. SETUP
1.1 Meter Zero
With no tubing connected to the device and without the “Push to Test” button
pushed, so that no power is applied to the pressure sensor, the needle
of the center-zero meter should lie over the zero mark of the scale,
in the
center of the window. If not, then the meter zero can be adjusted by
rotating the slotted adjuster below the needle. This is a delicate
adjustment and
should not be attempted unless necessary, and then only with care.
1.2 Electronics Zero
Still with no tubing attached to the device, push the “Push to
Test” button.
The LED will light up. There should be no movement of the needle
in the meter. If significant movement is observed, a potentiometer, accessible
through
a small hole in the top side of the device, may be adjusted to bring
the needle back to zero. Adjustment should only be made when power
is
applied
and the LED is lit. A small, flat-blade screwdriver is required to
adjust the potentiometer.
It is important when adjusting the electronics zero that no tubing
be attached to the device. This is because some sources of air
may have their own pressure
that is different from ambient and even a small difference would
be sufficient to apply a pressure difference across the pressure
sensor thus producing
a signal that would deflect the needle. To ensure that all pressures
in the device are equal to ambient, it is necessary to remove all
external tubing
from the device before adjusting the electronics zero.
1.3 Inlet filters
To protect the capillary tubes inside the device, inlet filters
are provided for both air inputs. Make sure that these filters
are securely
in place,
with the Luer connections properly mating to prevent any leakage.
The inlet filters fit directly into the range extender, between
the range extender
and the hose connectors.
1.4 Fresh air connection
Fresh air is used to dilute the high radon sample air. It should
have less than 0.5% of the sample air radon concentration for
no correction for the
radon in the fresh air to be needed. The upper limit of the
linear range of the RAD7 exceeds 10,000 pCi/L (~400,000 Bq/m3) so typically
the device
will not be used for radon concentrations in the sample air
significantly
less than 1,000 pCi/L, or 40,000 Bq/m3. 0.5% of that level
will be much greater than typical ambient radon concentration, so
for normal
purposes, when diluting
high radon concentration sample air, ambient air will be adequate
for the fresh air source.
On a still day there will be a significant vertical gradient
in radon concentration near the ground, out in the open.
To be confident
of
a low (<0.25 pCi/L
or <10 Bq/m3) ambient radon concentration it would be good to have
the fresh air sampling point 10 ft (3m) or more above the ground. If
the measurement
is being made inside a building, the fresh air sampling point should,
preferably, be outside the building and as high as conveniently possible.
If it is
a windy location, care should be taken to ensure that there are no
big pressure
changes in the supply due to shifts in the wind. In or on a wall facing
the wind that experiences the full Pitot pressure (static plus dynamic
pressure),
for example, would not be very good.
Please note that in some parts of the world the ambient radon
levels may be quite high and perhaps none of the precautions
listed above
will ensure
an insignificant radon concentration in the diluting air.
Also, if the device is used to dilute lower radon concentration
air,
even
normal ambient air
may not be sufficiently free of radon for its radon concentration
to be insignificant. In either of these circumstances,
for an accurate calculation of the original
high radon concentration, the radon in the diluting air
must be removed with an activated charcoal filter or measured
and a correction
made,
see below.
Plastic tubing from the fresh air sampling point to the
fresh air inlet filter on the device is all the fresh
air connection
required.
1.5 High-radon air connection
Operation of the range extender with some accessories,
such as a soil gas probe or a RAD AQUA radon-in-water
accessory, need
special consideration
and are covered later in individual chapters devoted
to their
cases. Other than that, in general, all that is needed
is a plastic tubing
connection
from the high-radon sampling point to the sample inlet
filter on the device.
The source should be at, or close to, ambient pressure.
If the source is not at ambient pressure, a pressure
reduction valve
and/or needle
valve may
be necessary. Alternatively, the fresh air pressure
could, perhaps, be changed to match the high radon
source pressure.
The point
is that the range extender
needle valve must be adjusted so that the pressure
drops across the capillary tubes in both air paths,
fresh and
sample,
are
always the
same (see below).
The extender uses a pressure drop to drive the air
through the capillary tubing in the device. The
RAD7 pump, therefore,
is
working against
resistance in the air path that reduces the total
air flow from around 750 ml/min to
about half that. For a ten to one reduction in
the RAD7 reading, nine tenths of the air flow will be
of fresh
air and only
one tenth will be of high-radon
sample air. Thus, the sample air flow will only
be about one ninth of the total flow, or about 50 ml/min
(0.05
L/min).
With a sample air flow rate of only 0.05 L/min,
it may take significant time for the sample to
reach
the RAD7.
For example,
if a 1 litre
drying unit is
interposed between the sampling point and the
range extender, it will take 20 minutes for the sample
to pass through
the drying unit. If response time
is of concern, therefore, the drying unit should
be placed between the range extender and the
RAD7. In
that case,
only one drying
unit is required as
all the incoming air, both fresh and high-radon
sample, has to pass through it on its way to
the RAD7.
If response time is not a concern, separate drying
units upstream of the Range Extender, one for
the high radon
sample air and
one for the fresh air,
would have the advantage that the air passing
through the capillary tubes is dry. This would
provide
additional protection
for
the capillary tubes
and prevent the growth of contamination.
If the high-radon sampling point is far from
the range extender and the RAD7, any effects
due to
the delay
in obtaining the
sample will
be ten times greater
than in making a measurement from the same
distance without the range extender. Whereas
it is acceptable,
for example,
to sample
soil gas
probes up to a
kilometer or more away with just the RAD7,
special operating procedures (see below)
must be adopted
to sample from
the same distance with
the range extender
in place.
1.6 RAD7 connection
From the ‘outlet to RAD7’ connection on the device, connect
tubing to the screw-cap end of a laboratory drying unit charged with
dry desiccant.
Connect the other end of the drying unit to the inlet filter on the
air inlet of the RAD7 (see RAD7 manual).
1.7 Pressure balancing
With the system set up, as above with connections
made to all three hose connectors, set
the pump to ON (Setup,
Pump,
ON
[ENTER]). You will hear
the pump operating. Press the “Push to Test” button on the Range
Extender and, while the LED is lit, adjust the needle valve until the meter
needle is centred on zero. If necessary repeat the process until the needle
stays zeroed when the “Push to Test” button is pressed.
If desired, once a rough zero is achieved, the high-sensitivity button
may
be pressed
to give a more sensitive indication of the pressure difference.
With the RAD7 pump not running, check
the pressure balance once more. If
there is
a pressure difference,
it means
that one or
the other
air source
is not at ambient pressure. This can
cause a problem as the excess pressure
may drive
air
through the
system even
when
the RAD7
pump is not running.
Once the pressure balancing is complete,
and if there is no source pressure
problem, set
the pump
to AUTO.
The Range
Extender
is
now set up, ready for
use.
If there is a source pressure problem,
then immediately prior to starting
the measurement
set the RAD7
pump to ON (Setup,
Pump, ON [ENTER]) so that
the pressure is balanced for the
duration of the measurement and
there are no intervals with the
pump off for the source pressure to push
a wrong
mixture through the system. After
the measurement is
complete, the pump should be reset
to AUTO to turn off the
pump.
The pressure balance may be checked
at any time during a measurement
(by pushing
the “Push to Test” button)
and the needle valve adjusted
if necessary.
2. Measurement
2.1 Operation
Once the Range Extender is set
up and the pump reset to
AUTO, the RAD7
may
be operated
in
whatever protocol
the
user wishes.
2.2 No DRYSTIK
Because of the significant
reduction in pressure between
the Range
Extender and
the RAD7 pump,
it is recommended
that a
DRYSTIK humidity
exchanger
not be used
in this setup. Not only would
it be less efficient, because
of the
low pressure,
but also the
inner membrane
tubing will
be more
likely to collapse
under the
suction generated by the RAD7
pump.
3. Calculation
3.1 Standard
With the fresh air radon concentration
less than 0.5% of the high-radon
sample air radon
concentration,
any
error
due to
radon in the fresh
air will be less
than 5% of the reading, and
may therefore be ignored.
The reading
on the RAD7
should be
multiplied by
the device multiplying factor,
as indicated on the Range
Extender itself.
If
A = radon concentration reading
M = multiplying factor
Then sample, high-radon concentration,
R, is given by
R = MA
3.2 Non-standard
If the diluting air has a
radon concentration more
than 0.5%
of the sample air,
then a correction has to
made as follows:
Let
A = radon concentration
reading
B = Fresh air radon concentration
R = Sample radon concentration
M = Multiplying factor
of Range Extender
In standard conditions,
with a multiplying factor
M, the
flow of high radon
air is 1/M of the total
flow
and the
flow of
fresh air
is (M – 1)/M
of the total flow.
So the air flowing to
the RAD7 will have
a radon
concentration of R/M
+ (M – 1)*B/M.
Now (M – 1)/M will be about 0.9 and, if B is small compared with R, (M – 1)/M
may be considered to be equal
to 1. Thus the air flowing to
the RAD7 will be nearly R/M +
B.
Therefore, before applying
the multiplying factor,
the fresh air
concentration,
B, should first
be deducted from A,
the RAD7 radon
concentration reading.
That is:
R = M(A – B)
4. Use with Soil Gas
Probe
4.1 Air flow
A typical sample
airflow rate into
the Range
Extender would
be around
0.05 L/min.
This
can be advantageous
in that
a continuous monitoring
of soil gas radon
concentration is
feasible. A 10 cm
radius sphere around
the sampling
point has a
volume of about
4L. At
0.05 L/min,
this would take
over an hour
to
draw. A 20
cm radius
sphere would take
10
hours to draw. So,
with the Range
Extender, it
is possible
to monitor
the radon
concentration
of soil
gas for many
hours,
thus getting a
very precise measurement,
without drawing so
much soil gas that
there is any danger
of fresh air reaching
the
sampling
point by
diffusion through
the soil.
4.2 Sample delay
On the other hand,
the very slow sample
flow
rate has
the disadvantage
that
it may take
considerable time
for the
soil gas to enter
the RAD7. If the
inside diameter
of the soil gas probe
is 0.5
cm and it is 1
m long, the
volume
will be about 20
ml. If the tubing
between
the probe and the
Range Extender
has an
inside diameter
of 3 mm
(about 1/8”),
the volume per metre
will be about 7 ml.
With two or three
metres of tubing,
the
total volume would
be less than
50 ml and the sample
would take less than
one minute to reach
the Range Extender
and sample
delay would not be
an issue.
If the soil gas
probe is some
distance from the
Range Extender,
then sample
delay may
be an issue.
For
example, with
the tubing above,
if the probe
is 1
km from the Range
Extender,
the
volume of the
tubing will be 7 L
and the sample
delay
would
be 140 min or
over 2
hours. Not
only
would this
require
waiting for three
hours to get
a reading, but also
the reading
obtained would
need correction
for
the decay
of
the radon in
the sample air
during the 2.3
hours it spent
traveling
along the tubing.
A strategy to
make it possible
to
sample from
far remote
probes without
having
to wait for
hours
for the sample
to arrive is
to use an external
pump to purge
the tubing
with sample air just
prior to
the measurement.
In other
words,
an external
pump
is attached
to the tubing and
run until
7 L has been
drawn from
the soil gas
probe. The
tubing is then connected
to the
Range Extender
and
a measurement
started.
Now the
RAD7
will be
drawing a valid
sample almost
immediately.
The measurement
can be continued
for
an hour
or more with
minimal loss
of
radon
from radioactive
decay
in the sampling
tube.
5. Use with
the RAD AQUA
5.1 Water
levels in
a standard
RAD AQUA.
In a standard
RAD AQUA
setup, all
the air
drawn from
the top hose
connection
is
returned
at
the flow
return in
the
side
of the RAD
AQUA. All
water has
some dissolved
gases in
it.
Usually the
concentration
of dissolved
gases
is slightly
more than
can be
supported
at equilibrium
at atmospheric
pressure.
So
there is
some out-gassing
during the
spray process
occurring
inside the
spray chamber.
As a result,
the water
level
in
the RAD AQUA
drops until
there
is a net
outflow
of air
through the
water release
channels
at the
bottom of
the spray
chamber.
In a rare
situation
of insufficient
dissolved
gases, so
that
the
water level
inside the
RAD AQUA
rises, a
bleed may
be connected
to
the ingoing
tube to allow
air
to
enter the
system to
replace that
lost in the
spray process.
This can
be
inserted
between
the return
check
valve and
the hose
connection
on the side
of
the RAD
AQUA.
5.2 Unequal
flows with
the Range
Extender.
With the
Range Extender
in
the circuit,
there
is ten
times more
air being
fed back
to the
RAD AQUA
than
being drawn
from it.
This is
not a serious
problem,
but the
air
coming
in is not
in equilibrium
with the
water
as it
has been
diluted
to one-tenth
the
equilibrium
concentration.
Much
of the
incoming
air
will be
blown out
of the
bottom
of
the RAD
AQUA, but
it will
take with
it some
air closer
to equilibrium
with
the water,
thus making
it harder
for
the RAD
AQUA to
keep the
air in
equilibrium
near the
top
of the
exchanger
from where
the sample
air is
drawn.
To limit
the air
returning
to the
RAD AQUA
to
the same
flow
as that leaving
from
the top hose
connection,
a simple
T-connection
with
a few
inches of tubing
connected
to it,
inserted
between
the
return
check
valve and
the RAD
AQUA
return hose
connection
will
provide a
passage
for the
excess
return
air to
escape.
There
will
then be
only
0.05 L/min airflow
through
the RAD
AQUA.
This will
give
plenty of time,
provided
the
water
flow rate is
more
than 0.5 L/min,
for the
returning
air to
recover
to
equilibrium
with
the water inside
the exchanger
before
the air
is once
more
on its way
to the
Range
Extender.
6. Maintenance
and
trouble shooting
6.1
Pressure
Balance
Should
it
ever
become
impossible
to
adjust
the
pressure
difference
to
zero,
with
the
needle
valve,
it
may
be
because
one
or
other
air
supply
is
at
a pressure
far
from
ambient.
To
check
this,
disconnect
both
air
supplies,
set
the
RAD7
pump
to
ON
and
attempt
to
reduce
the
pressure
difference
to
zero
by
adjusting
the
needle
valve.
If
a
zero setting,
with
supplies
disconnected,
is
now
possible,
look
for
reasons
why
one
or
other
supply
is
not
at
ambient
pressure
and
try
to
fix
the
problem.
Reconnect
the
supplies
and
observe
the
pressure
difference
with
the
RAD7
pump
not
running.
If
the input
pressure of
the two
air supplies
cannot be
changed, it
may be
possible to
fix the
problem by
stacking additional
inlet filters
on the
higher-pressure supply
to the
Range Extender,
thus creating
an additional
pressure drop
as the
airflow enters
the Range
Extender. In
this case,
as in
every case
where the
source pressures
are not
ambient, the
measurement should
be made
with the
RAD7 pump
running continuously
(see above).
If
a zero
balance is
impossible even
with the
supplies disconnected,
there is
a problem
with the
Range Extender.
Most likely
it will
be due
to a
dirty inlet
filter. Therefore,
replace both
inlet filters
and try
again.
If
this still
does not
solve the
problem, there
may be
some blockage
of a
capillary tube
inside the
Range Extender.
Provided the
Range Extender
is never
used without
the supplied
inlet filters
in place,
it should
never be
blocked by
any contamination.
However, if
blockage occurs,
the device
has to
be returned
to DURRIDGE
Company for
repair and
recalibration.
6.2
Batteries
The
Range Extender
uses 4
AA cells,
provided. The
electronics when
not in
use draw
a microscopic
current that
will discharge
the batteries
over 6
months to
a year.
The current
draw during
a reading,
after pushing
the “Push to Test” button,
is
small.
So,
in
regular
use,
the
batteries
can
be
expected
to
last
several
months.
When
the
batteries
need
replacing,
observed
by
a dim
panel
light
and
unresponsive
meter,
the
battery
compartment
on
the
faceplate
may
be
removed
and
the
batteries replaced. |